Date: Sun, 16 Mar 2003 11:57:47 -0500
All,
I got a chance yesterday to run some tests on my instrument.
Here's what I did:
I opened the loop and drove the force balance coil directly with a HP
function generator ( Model 3310B). I placed a 1000 ohm resistor in series
with the coil. I took my input signal from across the coil so I had a
sensitivity of 1000 volts/amp. The output was taken at the output of the
cap gauge circuit, after a 2 pole 35 Hz filter that would have been
inconvenient to remove. The output device was a HP 54603B memory scope. The
driven mass is pretty close to 500 grams.
I attempted to take phase and P-P position data at constant acceleration
from 0.1 Hz to 10 Hz while holding the current through the coil at 94
microamps, i. e. 94 millivolts across 1000 ohms. Why that value? Because it
was the maximum that I could use at 0.1 Hz and still maintain linearity of
the cap gauge signal.
Results:
The force balance coil/magnet combination have a transfer function of 0.26
Newtons/Amp.
I lost the signal in the noise at about 6 Hz. Obviously, I could go higher
with a larger input signal.
The phase between acceleration and displacement was less than 20 degrees
throughout the range.
The P-P displacement was about 60 nanometers at 6 Hz.
The laws of physics still apply, as the P-P position signal was inversely
proportional to the square of the frequency.
There was no sign of roll-off due to damping from the cap gauge, but then
the amplitudes were pretty low at the higher frequencies.
Comment: My low pass filters are set to roll off at 5 Hz so data beyond
that point is academic.
Dave ...